Document Type

Thesis

Date of Award

9-30-1983

Degree Name

Master of Science in Electrical Engineering - (M.S.)

Department

Electrical Engineering

First Advisor

Andrew Ulrich Meyer

Second Advisor

Stanley S. Reisman

Abstract

Electro-Oculography (EOG) is used in clinical ophthalmology to measure the time response of the corneoretinal potential (CRP) to changes in light intensity. An automatized data acquisition and system modeling link has been developed that significantly reduces processing time presently required for EOG. The data acquisition is performed on a microprocessor based instrument, which supervises the EOG-test, measures the potential and, after completion of the test, transfers the results to a mainframe computer. Parameter identification routines provide graphical representation and mathematical models for the CRP. Both nonlinear curve fitting methods and Least Squares System Identification schemes are used.