Error measurement for a portable tonometer
Document Type
Conference Proceeding
Publication Date
1-1-2002
Abstract
Most patients with glaucoma exhibit elevated intra-ocular pressure (IOP). A cost effective device that allows the patient to measure IOP at regular intervals may reduce vision loss. This research tested the Bausch & Lomb Proview self tonometer against a calibrated force meter. It further compared the Proview tonometer with results from the clinical gold standard, the Goldmann tonometer. Results show that the Proview tonometer has inaccuracies in calibration and uncertainties in clinical applications. If the Proview's calibration is corrected (using our results, e.g.), and the inaccuracies can be reduced, a device like the Proview has the potential to be clinically useful.
Identifier
11144345023 (Scopus)
Publication Title
Proceedings of the IEEE Annual Northeast Bioengineering Conference Nebec
e-ISSN
21607001
ISSN
1071121X
First Page
85
Last Page
86
Recommended Citation
Marchetto, Peter; Alvarez, Tara; Greene, Richard; and Thomas, Gordon, "Error measurement for a portable tonometer" (2002). Faculty Publications. 14758.