Ten-year evaluation of cementless Buechel-Pappas meniscal bearing total ankle replacement
Document Type
Article
Publication Date
6-1-2003
Abstract
A porous-coated, cementless, congruent-contact, three-piece, meniscal-bearing total ankle replacement was developed and used clinically over a 2- to 10-year period for patients with disabling ankle arthritis. Polished titanium-nitride ceramic-coated Ti6Al4V tibial and talar components with a deep-sulcus trochlear groove and two lateral fixation fins for the talar onlay component were used. The ultra-high-molecular-weight polyethylene (UHMWPe) meniscal bearing congruently conformed to the flat upper tibial component surface and the deep sulcus and cylindrical geometry of the lower talar component surface. Fifty deep-sulcus (Buechel-Pappas™) total ankle replacements were implanted in 49 patients. Diagnoses were 8 osteoarthritis (16%), 7 rheumatoid arthritis (14%), 2 avascular necrosis (4%), and 33 post-traumatic arthritis (66%). Ages ranged from 26 to 71 years (mean 49 years). Clinical results using a strict ankle scoring system demonstrated good/excellent results in 88% of cases. Postoperative ankle motion ranged from 12° to 46° total arc (mean 28°), which was similar to the preoperative motion. Revision for malalignment was necessary in two cases (4%). Mechanical complications included one case of meniscal bearing wear (2%) in a patient with posttraumatic arthritis with component malalignment and one case of talar component subsidence (2%) in a patient with avascular necrosis of the talus. No tibial component loosening was seen. Cumulative survivorship using an end point of revision of any component for any reason was 93.5% at 10 years (confidence interval 61 -100%).
Identifier
0038647750 (Scopus)
Publication Title
Foot and Ankle International
External Full Text Location
https://doi.org/10.1177/107110070302400603
ISSN
10711007
PubMed ID
12854666
First Page
462
Last Page
472
Issue
6
Volume
24
Recommended Citation
Buechel, Frederick F.; Buechel, Frederick F.; and Pappas, Michael J., "Ten-year evaluation of cementless Buechel-Pappas meniscal bearing total ankle replacement" (2003). Faculty Publications. 14111.
https://digitalcommons.njit.edu/fac_pubs/14111