Concise matrix analysis of point-based prostate targeting for intensity modulated radiation therapy
Document Type
Article
Publication Date
1-1-2007
Abstract
Intensity Modulated Radiation Therapy (IMRT) has recently emerged as an effective clinical treatment tool to treat various types of cancers by limiting the external beam dose to the surrounding normal tissue. However, the process of limiting external radiation dose to the tissue surrounding the tumor volume is not a trivial task. Several parameters including tumor volume and inhomogeneity, position and shape of the tumor volume, and the geometrical distribution of the radiation beams directly affect the determination of the external radiation dose. In addition, a major variable in effective delivery of the radiation dose is "set-up error" caused by the changes in patient position. Any changes in the position of the patient affect the geometrical location of the tumor volume and, therefore, need to be accommodated in the delivery of radiation beams during the treatment. This work presents a complete matrix representation required to calculate the three-dimensional rigid body homogeneous transformation matrices corresponding to external beam radiotherapy setup error and subsequent corrections in treatment beam parameters. A new concise orthogonal rotation solution is presented for use with clinical noisy data. Monte Carlo simulations prove the new matrix results are consistently better than the standard inverse solution. The required corrections in beam table, gantry, and collimator angles as function of the planned beam gantry angle are derived. For transformations that include a rotation on the sagittal plane, the required off-sets to beam parameters are complex functions of the planned gantry angle but are clearly documented graphically for clinical use. A case study is presented with an error analysis that supports the use of the presented method in a clinical environment. Clinical implementation and evaluation of the presented method with patient data is also included in the paper. ©Adenine Press (2007).
Identifier
33847756766 (Scopus)
Publication Title
Technology in Cancer Research and Treatment
External Full Text Location
https://doi.org/10.1177/153303460700600101
ISSN
15330346
PubMed ID
17241094
First Page
1
Last Page
10
Issue
1
Volume
6
Recommended Citation
Barbiere, Joseph; Hanley, Joseph; Song, Yulin; Dhawan, Atam P.; and Chan, Maria F., "Concise matrix analysis of point-based prostate targeting for intensity modulated radiation therapy" (2007). Faculty Publications. 13744.
https://digitalcommons.njit.edu/fac_pubs/13744